TM 10-4610-241-24
TO 40W4-13-42
3-36
CABLE ASSEMBLY, W57 (DISSOLVED SOLIDS SENSOR) REPAIR - continued.
b.
Cleaning.
Clean cable assembly and mounting hardware with damp rags.
c.
Inspection.
For inspection procedures, refer to TM 43-0158/TO 1-1A-15 (Air Force).
d.
Test
Check cable assembly as follows:
Check point A
Check point B
Required Value in Ohms
Pin B, J11
Outer Electrode
0.02
Pin A, J11
Inner Electrode
0.02
Pin C
Pin D
8 - 10K
Pin E
All other pins
Infinity
e.
Repair.
1.
Replace lockwashers and other parts if unserviceable.
2.
For repair procedures, refer to TM 43-0158/TO 1-lA-15 (Air Force).
f.
Installation.
(1)
Position cable assembly (17) and gasket (12) on back of control panel and install dust cap (10) four screws
(9), washers (8), lockwashers (7), nuts (6) and dust cap (10,).
(2)
Position two clamps (16) on back of control panel and secure with two screws (15), lockwashers (14) and
nuts (13).
NOTE
Be sure to wrap tape in the direction of threads.
(3)
If removed, install adapter (18) on pipe section (3).
(4)
Position gasket (4) and packing (5) on sensor (2) and install sensor on adapter (18). Tighten nut (1).
3-204