Table 2
Model 924 Functional Modes
FRONT PANEL SETTING
OUTPUTS(2)
PARAMETER 1 (3)
PARAMETER 2 (3)
FUNCTION
DISPLAY
RANGE
FRONT PANEL
TERM 1-10V FS
TERM 2-10V FS
SWITCH
SWITCH
SWITCH
DISPLAY
TERM 5-10mVFS
TERM 6-10mVFS
FAC
C1
10
10ppm HOCL FS
For all Front
5
HOCL in ppm
5ppm HOCL FS
Panel settings
1
1ppm HOCL FS
this parameter
gives pH from
0-14 FS.
TEMP
10
10ppm HOCL FS
5
Temp in deg-C
5ppm HOCL FS
1
1ppm HOCL FS
pH
10
10ppm HOCL FS
5
pH Sensor out-
5ppm HIOCL FS
1
put
1ppm HOCL FS
TFC/AUTO
CL
10
10ppm TFC FS
.(pH SENSOR)
5
TFC in ppm
5ppm TFC FS
1
1ppm TFC FS
TEMP
10
10ppm TFC FS
5
Temp. in C
5ppm TFC FS
1
1ppm TFC FS
pH
10
10ppm TFC FS
5
pH Sensor out-
5ppm TFC FS
1
put
1ppm TFC FS
TFC/MAN
CL
10
10ppm TFC FS
(pH FIXED)
5
TFC in ppm
5ppm TFC FS
1
1ppm TFC FS
pH(1)
10
10ppm TFC FS
5
Current fixed
5ppm TFC FS
1
pH
1ppm TFC FS
NOTES:
1)
When the FUNCTION switch is set to TFC/MAN and the DISPLAY switch to pH, the
instrument will be in the "pH SET-UP" mode. In this mode all calculations are suspended
to allow pH to be entered manually. Parameter 1 output is the value of TFC immediately
prior to switching to this setting. Parameter 2 is the current value of the pH setting.. The
data base is initialized when leaving this setting.
2)
Temperature indication is always available at output terminal 3, 0-10V representing 0-50C.
3)
Common for all voltage outputs is output terminal 7.
21