TM 10-6630-222-12&P
0004 00
OPERATOR AND UNIT MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST)
FOR WATER QUALITY ANALYSIS SET: PURIFICATION (WQAS-P)
DESCRIPTION AND USE OF CONTROLS AND INDICATORS FOR ULTRAMETER 6P
Press <CAL/MCLR> to enter calibration mode. The "CAL", "BUFFER" and "7" annunciators will
appear (see figure to the right). Displayed value will be the uncalibrated sensor.
CAL
7.00
NOTES: If a wrong buffer is added (outside of 6-8 pH), "7" and "BUFFER" will flash,
S
and the Ultrameter will not adjust. The uncalibrated pH value displayed in Step 4 will
7
BUFFER
assist in determining the accuracy of the pH sensor. If the pH reading is above 8 with pH
pH
7 buffer solution, the sensor well needs additional rinsing or the pH sensor is defective
and needs to be replaced.
/MS> or < /MR> until the display reads 7.0.
Press <
NOTE: Attempted calibration of >1 pH point from factory calibration will cause "FAC" to appear. This indicates the
need for sensor replacement or fresh buffer solution. The "FAC" internal electronic calibration is not intended to
replace calibration with pH buffers. It assumes an ideal pH sensor. Each "FAC" indicates a factory setting for that
calibration step (i.e., 7, acid, base).
You can press <CAL/MCLR> to accept the preset factory value, or you can reduce your variation from factory setting
by pressing < /MS> or < /MR>.
Press <CAL/MCLR> to accept the new value. The pH Zero Calibration is now complete. You may
exit by pressing any measurement key.
CHANGING FROM CENTIGRADE TO FAHRENHEIT
1.
Press <COND>
/MR> to display the stored memory records.
2.
Press <
Press < /MR> repeatedly until you pass the memory "clr ALL" location. The display will
3.
show a "C" or "F". (See figure to the right).
4.
Press <CAL/MCLR>, the display
will change to the other unit.
C
F
5.
Press <COND>, all temperature
readings are now in degrees last shown.
NOTE: Units symbol "F" is not displayed in Fahrenheit mode. Tempco will still be shown in %/C.
TOTAL RETURN to FACTORY SETTINGS
"FAC SEL"
There may come a time when it would be desirable to quickly reset all the recorded calibration values in the instrument
back to the factory settings. This might be to ensure all calibrations are set to a known value, or to give the instrument to
someone else free of adjustments or recorded data for a particular application.
1.
Press <COND>.
/MR> to display the stored memory records.
2.
Press <
0004 00-8