TM 10-6630-222-12&P
0006 00
OPERATOR AND UNIT MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST)
FOR WATER QUALITY ANALYSIS SET: PURIFICATION (WQAS-P)
DESCRIPTION AND USE OF CONTROLS AND INDICATORS FOR CHLORINE PHOTOMETER
ADD REAGENT TO SAMPLE (NOTE: The Powder Pop Dispenser (PPD) is used to add reagent to the sample water.)
A.
Invert the PPD to ensure the proper amount of powder is used.
B.
Remove the optical top and cuvette from the optical well.
C.
Remove the cap from the bottom of the PPD and from the top of the cuvette.
D.
Place the PPD over the top of the cuvette and fully depress the button on the top of the PPD to
dispense the reagent; replace cuvette and PPD covers.
E.
Invert cuvette a minimum of three (3) times to mix solution.
MEASUREMENT
A.
Place cuvette in optical well, ensuring clear portions are on left and right sides.
B.
Place black optical cover over cuvette.
C.
Immediately depress the "READ" pad.
(Measurements are in milligrams per liter [mg/l], [mg/l=ppm]
END OF WORK PACKAGE
0006 00-2