TM 10-6630-222-12&P
0007 00
OPERATOR AND UNIT MAINTENANCE MANUAL
(INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST)
FOR WATER QUALITY ANALYSIS SET: PURIFICATION (WQAS-P)
OPERATION UNDER USUAL CONDITIONS
turbidity of water and waste-water. The meter covers a 0-1000 FTU range in two scales: 0.00 to 50.00 FTU and 50 to
1000 FTU. The auto-ranging feature of the instrument sets the appropriate range for the measurement.
A. Turn the meter on by pressing the ON/OFF key.
B. The meter will carry out a self-test displaying a full set of figures. After the test, the LCD will change to the
measurement mode.
C. When the LCD displays "----" the meter is ready to measure.
D. Fill a clean cuvette up to one-quarter inch (0.5 cm) from its rim with the thoroughly agitated
sample. Allow
sufficient time for bubbles to escape before securing the cap.
Note
Do not over tighten the cap.
E. Wipe the cuvette thoroughly with a lint-free tissue before inserting into the measurement cell. The cuvette
must be completely free of fingerprints and other oil or dirt, particularly in the area where the light goes
through (approximately the bottom 2 cm/1 inch of the cuvette).
F.
Place the cuvette into the cell and check that the notch on the cap is positioned securely into the groove.
G. The mark on the cuvette cap should point towards the LCD.
H. Press the READ/⇑ key and the LCD will display a blinking "SIP" (Sampling in Process). The turbidity value
will appear after approximately 25 seconds.
CHLORINE PHOTOMETER
1.
PRINCIPLES OF OPERATION/ZEROING THE INSTRUMENT
A.
Turn the power on by depressing "ON/OFF" pad.
B.
Rinse cuvette a minimum of three (3) times with sample water.
C.
Fill the cuvette a maximum of inch from the top. DO NOT OVERFILL. Ensure excess water is
removed from the cuvette's exterior.
D.
Place the cuvette in optical well, ensuring clear portions are on left and right sides.
E.
Place black optical cover over cuvette.
F.
Depress "ZERO" pad.
ADD REAGENT TO SAMPLE (NOTE: The PPD is used to add reagent to the sample water.)
2.
A.
Invert the PPD to ensure the proper amount of powder is used.
B.
Remove the optical top and cuvette from the optical well.
C.
Remove the cap from the bottom of the PPD and from the top of the cuvette.
D.
Place the PPD over the top of the cuvette and fully depress the button on the top of the PPD to
dispense the reagent; replace cuvette and PPD covers.
E.
Invert cuvette a minimum of three (3) times to mix solution.
3.
MEASUREMENT
A.
Place cuvette in optical well, ensuring clear portions are on left and right sides.
B.
Place black optical cover over cuvette.
C.
Immediately depress the "READ" pad. (Measurements are in milligrams per liter [mg/l], [mg/l=ppm]
END OF WORK PACKAGE
0007 00-2