TM 10-6630-245-13&P
Table 2-1. Preventive Maintenance Checks and Services for WQAS-PM - Continued.
Location
Item
Item To Be
No.
Interval
Check/Service
Procedure
Not Fully Mission Capable If:
6
After
Test Kit, Dissolved
g. Inspect Thiosulfate Standard Solution
g. Bottle less than 25% full. Shelf life
Oxygen
for quantity, current shelf life, leaking
expired. Leakage is evident (until
(continued)
container, readable label.
cleaned up and replaced). Contents
unidentifiable.
7
After
Test Kit, Zinc
a. Inspect case and foam for function.
a. Use or environmental exposure has
made case or contents unreliable.
b. Inspect color comparator, two
b. Labware is broken or missing to the
measuring spoons, two comparison
extent that you can't make accurate
tubes, and pipet for cracks, breakage,
measurements.
damaged measurement markings. Make
sure labware is clean and dry.
c. Inspect Zinc Conditioning Reagent for
c. Bottle less than 25% full. Shelf life
quantity, current shelf life, leaking
expired. Leakage is evident (until
container, readable label.
cleaned up and replaced). Contents
unidentifiable.
d. Inspect Zinc Reagent for quantity,
d. Bottle less than 25% full. Shelf life
current shelf life, leaking container,
expired. Leakage is evident (until
readable label.
cleaned up and replaced). Contents
unidentifiable.
e. Inspect Deionized Water for quantity,
e. Less than one full bottle. Shelf life
current shelf life, leaking container,
expired. Leakage is evident (until
readable label.
cleaned up and replaced). Contents
unidentifiable.
8
After
Spectrophotometer a. Inspect case and fittings for function.
a. Use or environmental exposure has
made equipment unreliable.
b. Make sure unit is clean and dry.
c. Make sure sample cells are clean and
dry.
d. Inspect meter for damage, function.
d. Meter not intact or meter not
operating.
e. Inspect pH wide range meter cards for
e. There is not at least one unmarred,
damage.
undamaged scale.
f. Inspect fluoride meter cards for
f. There is not at least one unmarred,
damage.
undamaged scale.
g. Inspect iron meter cards for damage.
g. There is not at least one unmarred,
undamaged scale.
h. Inspect nitrogen ammonia meter cards
h. There is not at least one unmarred,
for damage.
undamaged scale.
i. Inspect nitrogen nitrate meter cards for
i. There is not at least one unmarred,
damage.
undamaged scale.
j. Inspect sulfate meter cards for damage.
j. There is not at least one unmarred,
undamaged scale.
2-8