Table 2-4. Calibration and Measurement Error Codes
Error Code
Description
Action/Remedy
E-20
Out of range
If this occurs only when electrode is out of
solution
being checked, the code should disappear
when returned to the solution
Check electrode and probe connections to
meter.
Recalibrate
using
fresh
standard
buffer
solutions.
E-21
Automatic calibration error
Press any key to reset.
Check that proper standard buffer solution is
being used.
Clean electrode bulb by vigorously stirring in
demineralized
or
distilled
water
or
by
squirting water from
wash bottle.
Recalibrate
using
fresh
standard
buffer
solution. If error
code repeats, replace electrode.
E-22
Calibration standard error. Two of
the same standard buffer solutions
are being measured.
Press any key to reset.
Check that two different standard buffer
solutions are
being used and the correct one is being
measured.
Recalibrate
using
fresh
standard
buffer
solutions.
E-23
Electrode slope is not within 80% to
120% range.
Press any key to reset.
Clean electrode bulb by vigorously stirring in
demineralized
or
distilled
water
or
by
squirting water from
wash bottle.
Recalibrate
using
fresh
standard
buffer
solution. If error
code repeats, replace electrode.
E-31
A value has been entered which is
not within the acceptable range.
Press any key to reset.
Enter a new value. Acceptable values are:
pH: -2.00 to +19.999.
pH Slope: 80% to 120%.
Timer Interval: 5 seconds to 23 hours 59
minutes.
2-31