LO BAT
This indicates that batteries are weak and have to be replaced. When this message appears, the meter
is able to perform 50 more measurements. See Battery Replacement step above.
- BA -
This indicates that batteries are too weak to perform measurements. The message appears for a few
seconds. Then the meter will automatically shut off.
ERR 1
This indicates that the cuvet is scratched or dirty or the calibration solutions are contaminated or
imprecise calibration procedure is used. If the message is still displayed after correcting these, the unit
may need to be replaced.