ARMY TM 10-4610-309-14
MARINE CORPS TM 10802A-14/1
0062 00
TACTICAL WATER PURIFICATION SYSTEM (TWPS)
UNIT MAINTENANCE PROCEDURES
CONTROL INSTRUMENTS
b. Press ENTER key to display Sensor A/Select Measure.
c. Press ENTER key again to display a screen Select Measure/(Conductivity ).
d. Use the Up or Down key to select desired measurement "Conductivity".
e. Press the ENTER key to enter the selected measurement.
3.
Select Display Format
a. With the Sensor A/Select Measure screen displayed, press the Down key once to
display Sensor A/Display Format.
b. Press the ENTER key to display a screen like Display Format/(200.0 us/cm).
c. Use the Down or Up keys to select a Display format .
1) Conductivity For Sensor A ENTER: 200.0 us/cm
2) Conductivity For Sensor B ENTER 2000 us/cm
d. Press ENTER key to enter the selected display format.
4.
Select Temperature Compensation
a. With the Sensor A/Display Format screen displayed, press the Down key once to
display Sensor A/T-Compensation.
b. Press ENTER key to display a screen like T-Compensation/Linear.
c. Use the Down and Up keys to select "LINEAR".
d. Then press the ENTER key to enter the selection.
5.
Select Config Linear
a. With the Sensor A/T-Compensation screen displayed press the Down key until Sensor
A/Config Linear screen appears.
b. Press ENTER key to display Config Linear/Set Slope.
c. Press ENTER key again to display a screen like Set Slope/(2.00 %/C ).
d. Use Arrow keys to adjust to a desired slope of "2.00 %/ C".
e. Press ENTER key to enter the slope.
f. After the Config Linear/Set Slope screen re-appears, press the Down key once to
display Config Linear/Set Ref Temp.
g. Press ENTER key to display a screen like Set Ref Temp/(25.0 C ).
h. Use the Arrow keys to adjust to the desired reference temperature of "25.0 C".
i. Press the ENTER key to enter the reference temperature.
j. After the Config Linear/Set Ref Temp screen re-appears, press ESC key once to return
to the Sensor A/Config Linear Screen.
6.
Select Cell K
a. With the Sensor A/Config Linear screen displayed, press the Down key until Sensor
A/Cell Constant screen appears.
b. Press ENTER key to display Cell Constant/Select Cell K.
c. Press ENTER key again to display a screen like Select Cell K/ (0.05000).
d. Use the Down and Up keys to select the nominal cell category that corresponds to the
sensor's GLI Certified "K" Value.
1) For Sensor A Enter "0.05000"
2) For Sensor B Enter "10"
e. Press ENTER key to enter the selected nominal cell category.
f. After the Cell Constant/Select Cell K screen re-appears, press the Down key once to
display Cell Constant/Set Cell K.
g. Press ENTER key to display a screen like Set Cell K/ (0.05000).
h. Use the Up or Down Arrow keys to adjust the display value to EXACTLY MATCH the
sensor's GLI certified "K" value listed on the Sensor cable.
i. Press ENTER key to enter the certified "K" value.
0062 00-15