TM 10-4610-310-14
WP 0049 00
OPERATOR MAINTENANCE
LIGHTWEIGHT WATER PURIFIER
ULTRAMETER
INITIAL SETUP
Materials/Parts
General Safety Instructions
Observe all CAUTIONS
GENERAL
This work package contains information and instructions for calibrating the Ultrameter instrument used on the
Lightweight Water Purifier (LWP).
Total Dissolved Solids (TDS)
1. Clean oily films or organic material from the cell electrodes with detergent. Do not scrub inside the cell.
2. Rinse out the cell with pure water after making measurements. Do not allow slow dissolving crystals to form
in the cell. This will contaminate future samples.
3. For maximum accuracy, keep the pH sensor cap on tight so no fluid washes into the conductivity cell.
CALIBRATION
Total dissolved solids (TDS), conductivity and pH are measured by using the Ultrameter from the BII box. See
Figure 1.
CONDUCTIVITY
PH SENSOR
CELL
CALIBRATION
BUTTON
Figure 1. Ultrameter
CALIBRATION continued
0049 00-1