CAUTION
The TDS meter is an accurate meter. With the exception of calibrating meter or replacing meter battery,
never attempt to service or modify instrument, nor allow it to get wet since moisture will damage the
meter.
a. Calibration.
Each time before using TDS meter, perform a
calibration check in
accordance with paragraph 2-
15c(4). Whenever switching to or
from seawater sources, meter must be
recalibrated in
accordance with paragraph 2-
(1)
Remove bottom plate.
(2)
Disconnect leads from battery.
(3)
Install new battery and connect leads.
(4)
Install bottom plate.
(5)
Depress indicator button and adjust wheel until the dial reading is e ual to the value stamped on the
q
(6) If reading cannot be adjusted after replacing battery, replace the TDS meter.
FIGURE 3-14. TOTAL DISSOLVED SOLIDS (TDS) METER BATTERY REPLACEMENT